KL-ST001
Portable Pressure Steam Sterilizer
- Dimension — 18 L590 × 505 × 450
- Dimension — 23 L640 × 505 × 450
The KL-ST005 horizontal steam sterilizer represents the large-capacity tier of Keling Medical's sterilization range, combining spacious chamber options with an intelligent control system designed for efficient, safe and consistent daily operation in demanding environments. Four volume options — 150 L to 360 L — allow central sterilization supply departments, pharmaceutical production lines and large research institutions to match the unit to their exact throughput without oversizing.
The stainless steel construction throughout the pressure vessel and chamber ensures corrosion resistance and structural integrity under the repeated thermal and pressure cycling of continuous hospital use. The intelligent control system provides real-time monitoring of temperature, pressure and cycle progress, with a programmable interface that simplifies operator interaction and reduces the likelihood of process deviations. Temperature uniformity ≤±2°C across all chamber sizes confirms that the thermal conditions reached every corner of the load, regardless of volume.
External steam supply (0.3–0.6 MPa) and vacuum capability (−0.086 MPa) enable pre-vacuum cycles that remove air pockets from porous loads before sterilization, supporting effective penetration into wrapped instrument sets and textile packs — an important requirement for surgical department use. Steam consumption is ≤14–25 kg and water consumption ≤190–200 kg depending on the chamber size selected.
150 L, 200 L, 250 L and 360 L options with matched overall dimensions for different facility throughputs.
real-time monitoring and programmable interface for accurate, operator-friendly cycle management.
covers pre-vacuum porous load cycles and standard steam sterilization protocols.
consistent thermal distribution across the full chamber confirms effective sterilization of every load item.
supports pre-vacuum air removal for effective steam penetration into wrapped and porous loads.
durable high-quality stainless steel pressure vessel for long service life under demanding daily use.
≤23 kW (150 L/200 L) and ≤35 kW (250 L/360 L) for fast heat-up and stable large-volume sterilization.
Specifications are subject to change; dimensional tolerances may apply. Contact Keling Medical for the latest datasheet.
| Model | KL-ST005 |
|---|---|
| Product type | Horizontal steam sterilizer |
| Volume (L) | 150 / 200 / 250 / 360 |
| Overall dimensions — 150 L | 950 × 1120 × 1760 mm |
| Overall dimensions — 200 L | 1175 × 1120 × 1760 mm |
| Overall dimensions — 250 L | 1030 × 1300 × 1900 mm |
| Overall dimensions — 360 L | 1250 × 1300 × 1900 mm |
| Chamber dimensions — 150 L | 700 × 450 × 480 mm |
| Chamber dimensions — 200 L | 925 × 450 × 480 mm |
| Chamber dimensions — 250 L | 780 × 600 × 600 mm |
| Chamber dimensions — 360 L | 1000 × 600 × 600 mm |
| Steam consumption — 150 L | ≤14 kg |
| Steam consumption — 200 L | ≤16 kg |
| Steam consumption — 250 L | ≤18 kg |
| Steam consumption — 360 L | ≤25 kg |
| Water consumption | ≤190–200 kg |
| Power — 150 L / 200 L | ≤23 kW |
| Power — 250 L / 360 L | ≤35 kW |
| Design pressure | −0.1 to 0.3 MPa |
| Temperature uniformity | ≤±2°C |
| Design temperature | 150°C |
| Water source pressure | 0.1–0.2 MPa |
| Working pressure | 0.23 MPa |
| Compressed air pressure | 0.4–0.7 MPa |
| Working temperature | 115–138°C |
| External steam pressure | 0.3–0.6 MPa |
| Vacuum | −0.086 MPa |
| Power supply voltage | Three-phase five-wire AC 380 V 50 Hz |
The KL-ST005 horizontal steam sterilizer is used in hospital central sterilization supply departments (CSSD), operating theater instrument services, pharmaceutical manufacturing sterilization suites and large clinical or research laboratories. Its pre-vacuum capability makes it especially suitable for wrapped surgical packs, textile loads, porous materials and complex instrument sets with lumens where direct steam penetration is critical.
The KL-ST005 is a high-capacity intelligent horizontal autoclave for sterilizing wrapped surgical instrument packs, textiles, dressings, culture media and porous loads in hospitals and pharmaceutical facilities. Its pre-vacuum capability (−0.086 MPa) ensures effective steam penetration even into complex wrapped loads and lumened instruments.
The KL-ST005 is available in 150 L (950 × 1120 × 1760 mm overall), 200 L (1175 × 1120 × 1760 mm), 250 L (1030 × 1300 × 1900 mm) and 360 L (1250 × 1300 × 1900 mm) configurations. Chamber dimensions range from 700 × 450 × 480 mm (150 L) to 1000 × 600 × 600 mm (360 L).
The intelligent control system offers real-time monitoring of temperature, pressure and cycle status with a programmable operator interface. This supports accurate cycle management, reduces the risk of human error and provides continuous process feedback throughout sterilization and drying phases.
Yes. Keling Medical supplies complete hospital sterilization projects and international distributors through its Shanghai export center. Contact Keling Medical at https://www.kelingmed.com for project specifications, installation requirements and export pricing.
Tell us the quantity, voltage and delivery port. The export team in Shanghai replies by email with pricing, lead time and the full datasheet.